AbstractPURPOSE:
In this study we systematically investigated different Dynamic Contrast Enhancement (DCE)-MRI protocols in the spine, with the goal of finding an optimal protocol that provides data suitable for quantitative pharmacokinetic modelling (PKM).
MATERIALS AND METHODS:
In 13 patients referred for MRI of the spine, DCE-MRI of the spine was performed with 2D and 3D MRI protocols on a 3T Philips Ingenuity MR system. A standard bolus of contrast agent (Dotarem - 0.2ml/kg body weight) was injected intravenously at a speed of 3ml/s. Different techniques for acceleration and motion compensation were tested: parallel imaging, partial-Fourier imaging and flow compensation. The quality of the DCE MRI images was scored on the basis of SNR, motion artefacts due to flow and respiration, signal enhancement, quality of the T1 map and of the arterial input function, and quality of pharmacokinetic model fitting to the extended Tofts model.
RESULTS:
Sagittal 3D sequences are to be preferred for PKM of the spine. Acceleration techniques were unsuccessful due to increased flow or motion artefacts. Motion compensating gradients failed to improve the DCE scans due to the longer echo time and the T2* decay which becomes more dominant and leads to signal loss, especially in the aorta. The quality scoring revealed that the best method was a conventional 3D gradient-echo acquisition without any acceleration or motion compensation technique. The priority in the choice of sequence parameters should be given to reducing echo time and keeping the dynamic temporal resolution below 5s. Increasing the number of acquisition, when possible, helps towards reducing flow artefacts. In our setting we achieved this with a sagittal 3D slab with 5 slices with a thickness of 4.5mm and two acquisitions.
CONCLUSION:
The proposed DCE protocol, encompassing the spine and the descending aorta, produces a realistic arterial input function and dynamic data suitable for PKM.
AuthorsCristina Lavini, Gem Kramer, Indra Pieters-den Bos, Otto Hoekstra, J T Marcus
JournalMagnetic resonance imaging (Magn Reson Imaging) Vol. 44 Pg. 96-103 (12 2017) ISSN: 1873-5894 [Electronic] Netherlands
PMID28867668 (Publication Type: Evaluation Study, Journal Article)
CopyrightCopyright © 2017 Elsevier Inc. All rights reserved.
Chemical References
  • Contrast Media
  • Organometallic Compounds
  • Meglumine
  • gadoterate meglumine
Topics
  • Algorithms
  • Artifacts
  • Contrast Media (pharmacokinetics)
  • Evaluation Studies as Topic
  • Female
  • Humans
  • Image Enhancement (methods)
  • Image Processing, Computer-Assisted (methods)
  • Imaging, Three-Dimensional (methods)
  • Magnetic Resonance Imaging (methods)
  • Male
  • Meglumine (pharmacokinetics)
  • Middle Aged
  • Organometallic Compounds (pharmacokinetics)
  • Sensitivity and Specificity
  • Spinal Diseases (diagnostic imaging)
  • Spine (diagnostic imaging)

Join CureHunter, for free Research Interface BASIC access!

Take advantage of free CureHunter research engine access to explore the best drug and treatment options for any disease. Find out why thousands of doctors, pharma researchers and patient activists around the world use CureHunter every day.
Realize the full power of the drug-disease research graph!


Choose Username:
Email:
Password:
Verify Password:
Enter Code Shown: