Analysis of the influence of 111In on 90Y-bremsstrahlung SPECT based on Monte Carlo simulation.

AbstractOBJECTIVE:
90Y-ibritumomab tiuxetan (Zevalin) which is used for the treatment of malignant lymphomas can be used for SPECT imaging based on bremsstrahlung from 90Y beta particles. However, gamma rays emitted by 111In, which is administered to evaluate the indication for the treatment, contaminate the 90Y bremsstrahlung images. Our objective is to investigate the influence of 111In on the 90Y SPECT images using Monte Carlo simulation.
METHODS:
We used an in-house developed simulation code for the Monte Carlo simulation of electrons and photons (MCEP). Two hot spheres with diameters of 40 mm were put in an elliptical phantom. Both spheres ("sphere 1" and "sphere 2") were filled with 90Y and 111In mixed solutions. The activities of 90Y in sphere 1 and sphere 2 were 241 and 394 kBq/mL, respectively, and the ones of 111In were 8.14 and 13.3 kBq/mL, respectively. The background activity of 90Y was 38.6 kBq/mL, whereas that of 111In was 1.30 kBq/mL; moreover, the acquisition time was 30 min. Two energy windows were used: one is 90-190 keV included the 111In photopeak; the other is 90-160 keV. To evaluate the quality of the SPECT images, the contrast recovery coefficient (CRC) and the constant noise ratio (CNR) of the SPECT images were derived.
RESULTS:
For the energy window between 90 and 160 keV, the 111In count was 74 % of the total. In that case, the CRC values were 30.1 and 30.7 % for "sphere 1" and "sphere 2", respectively, whereas the CNR values were 6.8 and 12.1, respectively. For the energy window between 90 and 190 keV, the 111In count reached 85 % of the total count. The CRC and CNR values were 38.6 and 40.0 % and 10.6 and 19.4, respectively.
CONCLUSIONS:
Our simulation study revealed that the cross talk between 111In and 90Y in SPECT imaging is rather serious. Even for the energy window excluding the 111In photopeak, the count ratio of 90Y was less than 30 % of the total. However, the influence of 111In on 90Y-SPECT imaging cannot be ignored, and the count ratio because of 111In is important to estimate the density of 90Y.
AuthorsHiroaki Shiba, Akihiko Takahashi, Shingo Baba, Kazuhiko Himuro, Yasuo Yamashita, Masayuki Sasaki
JournalAnnals of nuclear medicine (Ann Nucl Med) Vol. 30 Issue 10 Pg. 675-681 (Dec 2016) ISSN: 1864-6433 [Electronic] Japan
PMID27510893 (Publication Type: Journal Article)
Chemical References
  • Indium Radioisotopes
  • Yttrium Radioisotopes
Topics
  • Indium Radioisotopes
  • Monte Carlo Method
  • Phantoms, Imaging
  • Tomography, Emission-Computed, Single-Photon
  • Yttrium Radioisotopes

Join CureHunter, for free Research Interface BASIC access!

Take advantage of free CureHunter research engine access to explore the best drug and treatment options for any disease. Find out why thousands of doctors, pharma researchers and patient activists around the world use CureHunter every day.
Realize the full power of the drug-disease research graph!


Choose Username:
Email:
Password:
Verify Password:
Enter Code Shown: