MR thermometry analysis of sonication accuracy and safety margin of volumetric MR imaging-guided high-intensity focused ultrasound ablation of symptomatic uterine fibroids.

AbstractPURPOSE:
To evaluate the accuracy of the size and location of the ablation zone produced by volumetric magnetic resonance (MR) imaging-guided high-intensity focused ultrasound ablation of uterine fibroids on the basis of MR thermometric analysis and to assess the effects of a feedback control technique.
MATERIALS AND METHODS:
This prospective study was approved by the institutional review board, and written informed consent was obtained. Thirty-three women with 38 uterine fibroids were treated with an MR imaging-guided high-intensity focused ultrasound system capable of volumetric feedback ablation. Size (diameter times length) and location (three-dimensional displacements) of each ablation zone induced by 527 sonications (with [n=471] and without [n=56] feedback) were analyzed according to the thermal dose obtained with MR thermometry. Prospectively defined acceptance ranges of targeting accuracy were ±5 mm in left-right (LR) and craniocaudal (CC) directions and ±12 mm in anteroposterior (AP) direction. Effects of feedback control in 8- and 12-mm treatment cells were evaluated by using a mixed model with repeated observations within patients.
RESULTS:
Overall mean sizes of ablation zones produced by 4-, 8-, 12-, and 16-mm treatment cells (with and without feedback) were 4.6 mm±1.4 (standard deviation)×4.4 mm±4.8 (n=13), 8.9 mm±1.9×20.2 mm±6.5 (n=248), 13.0 mm±1.2×29.1 mm±5.6 (n=234), and 18.1 mm±1.4×38.2 mm±7.6 (n=32), respectively. Targeting accuracy values (displacements in absolute values) were 0.9 mm±0.7, 1.2 mm±0.9, and 2.8 mm±2.2 in LR, CC, and AP directions, respectively. Of 527 sonications, 99.8% (526 of 527) were within acceptance ranges. Feedback control had no statistically significant effect on targeting accuracy or ablation zone size. However, variations in ablation zone size were smaller in the feedback control group.
CONCLUSION:
Sonication accuracy of volumetric MR imaging-guided high-intensity focused ultrasound ablation of uterine fibroids appears clinically acceptable and may be further improved by feedback control to produce more consistent ablation zones.
AuthorsYoung-sun Kim, Hervé Trillaud, Hyunchul Rhim, Hyo K Lim, Willem Mali, Marianne Voogt, Jörg Barkhausen, Thomas Eckey, Max O Köhler, Bilgin Keserci, Charles Mougenot, Shunmugavelu D Sokka, Jouko Soini, Heikki J Nieminen
JournalRadiology (Radiology) Vol. 265 Issue 2 Pg. 627-37 (Nov 2012) ISSN: 1527-1315 [Electronic] United States
PMID23012465 (Publication Type: Clinical Trial, Journal Article, Research Support, Non-U.S. Gov't)
Copyright© RSNA, 2012
Topics
  • Adult
  • Female
  • High-Intensity Focused Ultrasound Ablation (methods)
  • Humans
  • Imaging, Three-Dimensional (methods)
  • Leiomyoma (diagnosis, surgery)
  • Magnetic Resonance Imaging (methods)
  • Middle Aged
  • Prognosis
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Surgery, Computer-Assisted (methods)
  • Thermography (methods)
  • Treatment Outcome
  • Uterine Neoplasms (diagnosis, surgery)

Join CureHunter, for free Research Interface BASIC access!

Take advantage of free CureHunter research engine access to explore the best drug and treatment options for any disease. Find out why thousands of doctors, pharma researchers and patient activists around the world use CureHunter every day.
Realize the full power of the drug-disease research graph!


Choose Username:
Email:
Password:
Verify Password:
Enter Code Shown: