Prospective evaluation of a new self-expanding plastic stent for inoperable esophageal strictures.

AbstractBACKGROUND:
High stent cost is considered the major drawback of self-expanding metal stents for dysphagia palliation in patients with inoperable esophageal strictures. We report our experience with a self-expanding plastic (Polyflex) stent, the cost of which is half that of the metal stents.
METHODS:
Between September 1999 and April 2001, 16 dysphagic patients (15 men; mean age, 69.4 +/- 14.5 years; range, 49-100 years; mean dysphagia score, 3.31 +/- 0.6) with esophageal strictures who underwent Polyflex stent placement (insertion device diameter, 12-14 mm; postexpansion inner stent diameter, 16-21 mm; stent length, 9, 12, and 15 cm) were studied prospectively. The strictures were caused by postsurgical recurrence of gastric/esophageal cancer at the anastomotic site in five patients, primary esophageal cancer in four patients, esophagocardia junction cancer in four patients, metastatic mediastinal lymph nodes from a primary lung cancer invading the esophagus in 1 patient, and benign peptic stricture in two elderly patients. All the patients were prospectively followed until death.
RESULTS:
Stent insertion was technically successful in 12 patients (75%). Stent placement failed in four patients (25%) because of failure to pass the delivery catheter across the stricture in three patients and failure of the stent to open in one patient. Early and late stent migration occurred in two patients and 1 patient, respectively. Tumor overgrowth occurred in 1 patient. The mean dysphagia score 7 days after stent placement was 1.1 +/- 0.9. Mean survival was 100.6 +/- 71.2 days (range, 8-225 days).
CONCLUSION:
Self-expanding Polyflex stents are safe and effective for inoperable esophageal strictures and have an acceptable technical success rate. Further experience, better selection criteria, and design improvements should improve results.
AuthorsG Costamagna, S K Shah, A Tringali, M Mutignani, V Perri, M E Riccioni
JournalSurgical endoscopy (Surg Endosc) Vol. 17 Issue 6 Pg. 891-5 (Jun 2003) ISSN: 1432-2218 [Electronic] Germany
PMID12618935 (Publication Type: Evaluation Study, Journal Article)
Chemical References
  • Plastics
Topics
  • Aged
  • Aged, 80 and over
  • Deglutition Disorders
  • Equipment Failure
  • Esophageal Stenosis (pathology, surgery, therapy)
  • Follow-Up Studies
  • Foreign-Body Migration
  • Humans
  • Male
  • Middle Aged
  • Plastics (therapeutic use)
  • Prospective Studies
  • Stents
  • Treatment Failure
  • Treatment Outcome

Join CureHunter, for free Research Interface BASIC access!

Take advantage of free CureHunter research engine access to explore the best drug and treatment options for any disease. Find out why thousands of doctors, pharma researchers and patient activists around the world use CureHunter every day.
Realize the full power of the drug-disease research graph!


Choose Username:
Email:
Password:
Verify Password:
Enter Code Shown: